Ci-120i is a fully automatic bench top analyzer consisting of an analytical module and a control module |
![]() |
Operates on whole blood or platelet- poor plasma | platelet-poor plasma |
platelet-poor plasma | racks, primary tubes |
Model type | benchtop |
Dimensions (H × W × D)- analyzer Dimensions (H × W × D)- Outer packing |
640mm×800mm ×630mm 860mm×950mm ×755mm |
Weight- analyzer Weight- Outer packing |
85kg 125kg |
Electric Ratings: | ac. 110-240V, 50/60Hz, <800VA |
Ambient Temperature: | Between 15℃ and 30℃ |
Relative Humidity: | Between 35% and 85 % |
Methodologies supported | clot detection (LED optical transmission method); chromogenic; immunologic; Wavelength 405nm;545nm;670nm;800nm |
Number of different measured assays onboard simultaneously | 15 |
Number of different assays programmed and calib. at one time | unlimited |
Factor assays require manual manipulation or dilutions | No |
Number of reagent containers onboard at once | 24 (reagent area,8℃) |
Multiple reagent configurations supported | Yes |
Walkaway capacity: Number of specimens/Number of tests | 25/up to 500 |
Minimum sample volume aspirated precisely at one time | 3ul |
Standard specimen volume required to run PT or PTT/Factor VIII activity | 50ul/25ul |
Primary tube sampling supported/Pierces caps on primary tubes | Yes/No |
Sample bar-code reading capability | Yes |
Reagent bar-code reading capability | Yes |
Onboard test automatic inventory | Yes |
Measures No. of tests remaining/Short sample detection | Yes/Yes |
Auto. detects adequate reagents for aspiration and analysis | Yes |
Dilution of patient samples onboard | Yes |
Automatic rerun capability/Auto reflex testing capability | Yes |
Read time extended for prolonged clotting times | Yes |
User can set different-than-standard | 1) Reagent volumes/Sample volumes; 2) No. and sources of reagent; 3) Incubating times/Reading times; |
Autocalib. Or autocalib. Alert / Multipoint calib. supported | Yes |
Stat time to complete all analyses/Throughput per hour for: | PT Alone:≦6 minutes/150 specimens PT, PTT:≦6 minutes/80 specimens D-dimer:≦6 minutes/80 specimens Factor VIII activity assay:≦11 minutes/60 specimens |
Time delay from ordering STAT to aspiration of sample | minimal |
Number of Measurements | At each standard point, up to 5 simultaneous multi-parameter analysis is available. |
Automatic transfer of QC results to LIS | Yes |
Protection Function | Over-heat Protection Fuse Mechanical stop switch Cover open/close sensor Probe protection sensor Over-current Protection Fuse |
Unit Error Alert Function |
Instrument Abnormality-Monitoring Function |
automatic supply reaction cuvettes | Yes |